Username: Password:

PWB43

Serial number: 43 item id: 11035
PO number:
Received:
Manufacturer:
Model:
Created on:
Updated on:
Notes:

close

Date User Note
2018-05-15 00:00:00 SFP not reporting diagnostics, no matter which SFP is plugged in
2018-06-21 00:00:00 problem with SFP persists
2019-07-22 00:00:00 working, as far as I can tell
2019-10-07 00:00:00 intermittent no boot. eth link ok, sata link ok, no ping. K.O.
2019-10-07 00:00:00 no boot. K.O.
2019-10-07 00:00:00 after reflash of epcq, boot ok, but no boot from user page. K.O.
2019-10-10 00:00:00 vcxo replaced (Luke), boot ok, ext clock ok. K.O.
2019-10-10 00:00:00 1 dead channel: seqpad 226, sca3 chan 47. K.O.
2019-10-10 00:00:00 found and fixed bent pin on the TPC connector corresponding to sca3 chan47. fixed the bent pin, channel is good now. K.O.
2019-10-10 00:00:00 all pads fw pulser signal ok. K.O.
2019-10-10 00:00:00 intermittent boot failure: - Running Low Frequency Pattern Test... Failure! Block Trail Distance: 0xff, Block size: 0x100000, Transfer Length: 0x20000000 - Comparing Flash Image to Loaded Image in DDR... !!Flash/DDR Mismatch at 0x20!! K.O.
2019-10-18 00:00:00 does not boot. K.O.
2019-10-30 00:00:00 always does not boot, memory test failure: Starting Memory Checker... - Running Walking Zero Pattern Test... and then, nothing. K.O.
2020-09-17 00:00:00 memory error: DDR CSR#4 reads 0x6 (pll_locked and calibration failure). quartus memory toolkit reports calibration errors: DQS group 0: Read Calibration - Per-bit read deskew failure. K.O.
2020-09-17 00:00:00 after gently tapping on the memory chips and the fpga chip, ddr csr at least once read 0x5 (pll_locked and calibration success), the firmware booted NIOS to DHCP. But quartus memory toolkit still reports same error for DQS group 0. K.O.
2020-10-02 00:00:00 sent to Dorigo to replace DDR chips. K.O.
2020-10-06 00:00:00 returned from Dorigo. ddr memory calibration error, dqs group 0 and 1 (first memory chip). K.O.