Date |
User |
Note |
2018-06-01 00:00:00 |
|
Clock didn't lock, replaced VCXO, now seems to work.
SCAs need testing still. |
2018-06-05 00:00:00 |
|
SCAs have wrong offset |
2018-06-06 00:00:00 |
|
resistors to fix |
2019-05-14 00:00:00 |
|
labeled "all SCA wrong offset" K.O. |
2019-09-23 00:00:00 |
|
boot ok, ping ok, esper ok, all sca ok, wrong offset all channels. K.O. |
2019-10-08 00:00:00 |
|
finished replacing baseline offset resistors. all offsets good now. K.O. |
2019-10-08 00:00:00 |
|
dead channels: seqpad 180, 181, 187, 252, 269. K.O. |
2019-10-08 00:00:00 |
|
dead channels: sca2 chan 37, 54, 65, 71, 72. K.O. |
2019-10-09 00:00:00 |
|
dead channels: sca2 chan 65. K.O. |
2019-10-10 00:00:00 |
|
replaced broken capacitor sca2 chan65 seqpad187, now pad completely dead. K.O. |
2019-10-13 00:00:00 |
|
flex cable is damaged. SCA C and SCA D read temperature -999. K.O.
|
2023-06-15 00:00:00 |
|
Removed from TPC |