Detector Calibration: Difference between revisions

From AgWiki
Jump to navigation Jump to search
Operation: starting write-up
Fixed references, removed spurious paste debris
 
(6 intermediate revisions by the same user not shown)
Line 3: Line 3:
= Table of Figures =
= Table of Figures =


[[#_Toc493167186|Figure 1 – rTPC cross-section with at the bottom the central cathode with Al strips. A laser plane illuminates the central cathode for electron production. 5]]
[[#_Fig_crosssec|Figure 1 – rTPC cross-section with at the bottom the central cathode with Al strips. A laser plane illuminates the central cathode for electron production]]


[[#_Toc493167187|Figure 2 - Inner cylinder with copper cathode and Aluminum strips 6]]
[[#_Fig_cathode|Figure 2 - Inner cylinder with copper cathode and Aluminum strips]]


[[#_Toc493167188|Figure 3 - Optical head assembly to be mounted on the endplate 6]]
[[#_Fig_spreader|Figure 3 - Optical head assembly to be mounted on the endplate]]


[[#_Toc493167189|Figure 4 - Model of the laser equipment at the endplate with light simulated plane on first Aluminum strip. 7]]
[[#_Fig_lightsim|Figure 4 - Model of the laser equipment at the endplate with light simulated plane on first Aluminum strip]]


[[#_Toc647854563|Figure 5 - Pulse height vs pad row from prototype]]
[[#_Fig_ph_v_row|Figure 5 - Pulse height vs pad row from prototype]]


[[#_Toc647854564|Figure 6 - Laser shutter positions]]
[[#_Fig_shutter|Figure 6 - Laser shutter positions]]


[[#_Toc647854565|Figure 7 - Interlock cabling]]
[[#_Fig_fiberbox|Figure 7 - Laser fiber connection box]]


[[#_Toc647854566|Figure 8 - MIDAS laser page]]
[[#_Fig_laserbox|Figure 8 - Laser box and MEMS driver]]


[[#_Toc647854567|Figure 9 - Laser trigger setup]]
[[#_Fig_midas|Figure 9 - MIDAS laser page]]


[[#_Toc493167190|Figure 10 - UBC setup for laser test]]
[[#_Fig_trigger|Figure 10 - Laser trigger setup]]


[[#_Toc493167191|Figure 11 - Schematic of the UBC test configuration]]
[[#_Fig_tpclasersig|Figure 11 - Test for laser light in TPC]]


[[#_Toc493167192|Figure 12 - Photoemission versus incident angle for Aluminum]]
[[#_Fig_ubcsetup|Figure 12 - UBC setup for laser test]]


[[#_Toc493167193|Figure 13 - Ratio of electron per photon versus incident angle for different base material]]
[[#_Fig_ubcschem|Figure 13 - Schematic of the UBC test configuration]]
 
[[#_Fig_ubcres1|Figure 14 - Photoemission versus incident angle for Aluminum]]
 
[[#_Fig_ubcres2|Figure 15 - Ratio of electron per photon versus incident angle for different base material]]


= Purpose =
= Purpose =
Line 53: Line 57:
The top and bottom endplates have provisions for inserting the necessary optic module to deliver a laser plane intersecting the inner cathode cylinder. The initial laser beam in the order of 0.5mm in diameter is converted into a plane through a transparent cylindrical quartz rod of 5mm diameter within the detector volume. The plane has an aperture angle in the order of 90deg illuminating both sides of the drift path. Aluminum strips on pre-defined locations along the inner cathode will emit electrons when struck by the 266nm laser light. Using a triggered pulse laser source, the electron arrival on the anode wire and the induced charge on the cathode pads along the anode wire will provide the timing and Z position of the intersection point of the aluminum strip and the laser plane. The light plane is slightly rotated and tilted off the z-axis in order separate the Al strip electron cloud from reaching all the same wires. For the given rotation and tilt angles (45deg, 1deg) we have spaced the Al strips such that 2 anode wires will be separating the 2 electron clouds of two consecutive Al strips.
The top and bottom endplates have provisions for inserting the necessary optic module to deliver a laser plane intersecting the inner cathode cylinder. The initial laser beam in the order of 0.5mm in diameter is converted into a plane through a transparent cylindrical quartz rod of 5mm diameter within the detector volume. The plane has an aperture angle in the order of 90deg illuminating both sides of the drift path. Aluminum strips on pre-defined locations along the inner cathode will emit electrons when struck by the 266nm laser light. Using a triggered pulse laser source, the electron arrival on the anode wire and the induced charge on the cathode pads along the anode wire will provide the timing and Z position of the intersection point of the aluminum strip and the laser plane. The light plane is slightly rotated and tilted off the z-axis in order separate the Al strip electron cloud from reaching all the same wires. For the given rotation and tilt angles (45deg, 1deg) we have spaced the Al strips such that 2 anode wires will be separating the 2 electron clouds of two consecutive Al strips.


<span id="_Fig_crosssec" class="anchor"></span>
[[File:imagec2.png|546x138px]]
[[File:imagec2.png|546x138px]]


<span id="_Ref450313614" class="anchor"><span id="_Toc493167186" class="anchor"></span></span>Figure 1 – rTPC cross-section with at the bottom the central cathode with Al strips. A laser plane illuminates the central cathode for electron production.
Figure 1 – rTPC cross-section with at the bottom the central cathode with Al strips. A laser plane illuminates the central cathode for electron production.


= Mechanical design =
= Mechanical design =
Line 61: Line 66:
The mechanical modification to the endplate for the laser calibration option is minimal. They consist in:
The mechanical modification to the endplate for the laser calibration option is minimal. They consist in:


* Placing 2 laser access holes in each endplate, giving us the option to feed 4 laser optic modules to the rTPC, see Figure 3. The advantage of multiple access points is to better cover the whole drift volume and to illuminate both ends of the detector with similar incident angles.
* Placing 2 laser access holes in each endplate, giving us the option to feed 4 laser optic modules to the rTPC (see [[#_Fig_spreader|Figure]]). The advantage of multiple access points is to better cover the whole drift volume and to illuminate both ends of the detector with similar incident angles.
* Additional 9 Aluminum strips of 6mm width wrapped around the inner cylinder at 9 locations see Figure 7. Table 1 list the position of the 9 strips, the final column shows the phi coordinate at which the laser light hits the given strip for one possible set of geometric parameters of the optics package.
* Additional 9 Aluminum strips of 6mm width wrapped around the inner cylinder at 9 locations (see [[#_Fig_cathode|Figure]]). [[#_Table1|Table 1]] list the position of the 9 strips, the final column shows the phi coordinate at which the laser light hits the given strip for one possible set of geometric parameters of the optics package.


<span id="_Table1" class="anchor"></span>
{|
{|
! Strip number
! Strip number
Line 105: Line 111:
| 34.348
| 34.348
|}
|}
Table 1 - Al strip position


<span id="_Ref490026846" class="anchor"><span id="_Ref490026835" class="anchor"></span></span>Table 1 - Al strip position
<span id="_Fig_cathode" class="anchor"></span>
 
[[File:imagec3.png|576x268px]]
[[File:imagec3.png|576x268px]]


<span id="_Toc493167187" class="anchor"></span>Figure 2 - Inner cylinder with copper cathode and Aluminum strips
Figure 2 - Inner cylinder with copper cathode and Aluminum strips


<span id="_Fig_spreader" class="anchor"></span>
{|
{|
| [[File:imagec4.jpeg|305x229px]]
| [[File:imagec4.jpeg|305x229px]]
| [[File:imagec5.jpeg|207x155px]][[File:imagec6.jpeg|210x157px]]
| [[File:imagec5.jpeg|207x155px]][[File:imagec6.jpeg|210x157px]]
|}
|}
Figure 3 -  Optical head assembly to be mounted on the endplate 6


<span id="_Fig_lightsim" class="anchor"></span>
[[File:imagec7.png|416x318px]]
[[File:imagec7.png|416x318px]]


<span id="_Ref479591070" class="anchor"><span id="_Toc493167189" class="anchor"></span></span>Figure 4 - Model of the laser equipment at the endplate with light simulated plane on first Aluminum strip.
Figure 4 - Model of the laser equipment at the endplate with light simulated plane on first Aluminum strip.


== Specifications ==
== Specifications ==
Line 172: Line 181:
Tests of the laser calibration system using the prototype show a clear pattern of the Al strips. Analysis of the pattern reproduces the strip position with good accuracy. Despite the strip pitch in the prototype being much tighter at 40mm instead of 265mm the peaks are very clearly separated in the z-pads. The amplitude variation between peaks is due to the laser intensity distribution which has not been investigated in detail. It is not a high priority because the laser system is not meant for pulse height calibration, but rather for time and position.
Tests of the laser calibration system using the prototype show a clear pattern of the Al strips. Analysis of the pattern reproduces the strip position with good accuracy. Despite the strip pitch in the prototype being much tighter at 40mm instead of 265mm the peaks are very clearly separated in the z-pads. The amplitude variation between peaks is due to the laser intensity distribution which has not been investigated in detail. It is not a high priority because the laser system is not meant for pulse height calibration, but rather for time and position.


<span id="_Fig_ph_v_row" class="anchor"></span>
[[File:imagec8.png|550x358px]]
[[File:imagec8.png|550x358px]]
<span id="_Ref567783653" class="anchor"><span id="_Toc647854563" class="anchor"></span></span>Figure 5 - Pulse height vs pad row from prototype
Figure 5 - Pulse height vs pad row from prototype


The laser plane is not inclined in the prototype, so a separation on the anode wires is impossible. The wider spacing for the final detector was chosen to allow for a good separation in the anodes.
The laser plane is not inclined in the prototype, so a separation on the anode wires is impossible. The wider spacing for the final detector was chosen to allow for a good separation in the anodes.
Line 179: Line 189:
As mentioned above the laser intensity distribution is not well understood yet, but we are confident we can illuminate the 4-5 strips closest to each laser port. The combination of shallow incidence angle and low geometric efficiency makes it questionable whether enough light can reach the far end of the TPC. It may be possible to increase the laser intensity to reach further strips at the expense of saturating the signals in the intensity maximum.
As mentioned above the laser intensity distribution is not well understood yet, but we are confident we can illuminate the 4-5 strips closest to each laser port. The combination of shallow incidence angle and low geometric efficiency makes it questionable whether enough light can reach the far end of the TPC. It may be possible to increase the laser intensity to reach further strips at the expense of saturating the signals in the intensity maximum.


== Operation ==
== Physical safety measures and Interlock ==
=== Physical safety measures and Interlock ===
=== Shutter and Box ===
The laser is equipped with a physical shutter, that prevents laser light from coming out even if the laser is firing. '''Ensure the shutter is closed before disconnecting fiber. Ensure it's open when trying to operate.'''
The laser is equipped with a physical shutter, that prevents laser light from coming out even if the laser is firing. '''Ensure the shutter is closed before disconnecting fiber. Ensure it's open when trying to operate.'''
<span id="_Fig_shutter" class="anchor"></span>
{|
{|
|
|
Line 188: Line 200:
[[File:Laser shutter open.png|thumb|Physical shutter open, laser can operate]]
[[File:Laser shutter open.png|thumb|Physical shutter open, laser can operate]]
|}
|}
Figure 6 - Laser shutter positions


<span id="_Ref567783654" class="anchor"><span id="_Toc647854564" class="anchor"></span></span>Figure 6 - Laser shutter positions
The laser head and multiplexer are> I noticed that for some reason the per-port Q-switch delay setting in the ODB was way higher for port T11 (elog:6574/1 [3], higher delay -> less laser power), and indeed setting it to a value more in line with the others (210) brings back the coincidence triggers, so we do in fact get laser light from all 4 ports.
> Running a longer sequence now, and will turn on PWBs later.


The laser head and multiplexer are enclosed in a locked metal case, which is hooked up to the interlock input on the laser power supply, which means the laser will not fire if the case is open. (As always with electronic interlocks, do not rely on this, but confirm the interlock light on the handheld control box is blinking when the box is open.)
enclosed in a locked metal case, which is hooked up to the interlock input on the laser power supply, which means the laser will not fire if the case is open. (As always with electronic interlocks, do not rely on this, but confirm the interlock light on the handheld control box is blinking when the box is open.)


The combination for the lock is the wavelength of the laser.
The combination for the lock is the wavelength of the laser.
Line 198: Line 212:


The laser fibers are enclosed in a steel conduit, with a small locked metal box protecting the interconnections between laser-side and TPC-side fibers.
The laser fibers are enclosed in a steel conduit, with a small locked metal box protecting the interconnections between laser-side and TPC-side fibers.
<span id="_Fig_fiberbox" class="anchor"></span>
{|
{|
|
|
Line 205: Line 221:
|
|
|}
|}
Figure 7 - Laser fiber connection box
== MEMS multiplexer ==
In addition to the laser head itself, the laser box contains a fiber multiplexer based around a micro-electromechanical system (MEMS) mirror from [https://www.mirrorcletech.com/ Mirrorcle Technologies]:
{| class="wikitable"
|-
! Device !! Part No. !! documentation
|-
| MEMS mirror  || A7M20.2-2000AL || [[ File:MirrorcleTech_Datasheet_A7M20.2-2000AL.pdf|datasheet ]]
|-
| Driver board || DR-10-056-00 || [[ File:Mirrorcle_MEMS_Drivers_5.x_-_User_Guide.pdf|User guide ]]
|-
| Breakout board || BRK-DRIVER-5.x||
|-
|}
The MEMS driver board is controlled via SPI from a raspberry pi, which also provides a required 30kHz clock and the 5V supply voltage it needs. The driver board generates about 200V steering voltage for the MEMS from this supply, so should not be touched when operational. When the driver board is connected correctly to a correctly configured raspberry pi, both the green and blue LED should be illuminated. The red LED is only lit when the high voltage is turned on, which is done via a GPIO pin from the pi.
<span id="_Fig_laserbox" class="anchor"></span>
{|
|
[[File:Laserbox.png|thumb|Laserbox interior with head (A), lens (B), MEMS mirror (C), MEMS driver (D), fiber port array (E)]]
|
[[File:MEMSdriver.png|thumb|MEMS driver board with controller connector (A, to breakout board, then rPi), status LEDs (B), voltage testpoints (C), HV connector to MEMS (D)]]
|}
Figure 8 - Laser box and MEMS driver board


== Operation ==
=== Manual operation / Switching on ===
=== Manual operation / Switching on ===
The laser gets turned on with the key at the front of the power supply unit. After turning on it may take a while (~5min) to warm up the cooling water (interlock message is displayed on the handheld control unit). Other interlock messages indicate something is wrong.
The laser gets turned on with the key at the front of the power supply unit. After turning on it may take a while (~5min) to warm up the cooling water (interlock message is displayed on the handheld control unit). Other interlock messages indicate something is wrong.


Once the system is started up it can be operated from the handheld control unit, by pressing first the flash start, then after a mandatory waiting period the QS start buttons. This manual operatio should not be necessary.
Once the system is started up it can be operated from the handheld control unit, by pressing first the flash start, then after a mandatory waiting period the QS start buttons. This manual operation should not be necessary.


=== Midas operation ===
=== Midas operation ===
The Midas laser page mirrors the important parts of the manual control. In order to use it the feLaser frontend must be running. To start firing the laser, press the "Start flash lamp" button, then wait until the status line shows "flash", then press "Start Q-Switch". Pausing the laser briefly can be done by just stopping the Q-Switch, for longer breaks turn off the flash lamp. (Turning off the flash lamp automatically turns off the Q-Switch, so there is no need to turn it off manually.
The Midas laser page mirrors the important parts of the manual control. In order to use it the feLaser frontend must be running. To start firing the laser, press the "Start flash lamp" button, then wait until the status line shows "flash", then press "Start Q-Switch". Pausing the laser briefly can be done by just stopping the Q-Switch, for longer breaks turn off the flash lamp. (Turning off the flash lamp automatically turns off the Q-Switch, so there is no need to turn it off manually.
<span id="_Fig_midas" class="anchor"></span>
{|
{|
|
|
[[File:Screenshot from 2018-11-08 14-55-42.png|thumb|Midas laser page: operate laser by starting flash lamp, then QS]]
[[File:laserpage.png|thumb|Midas laser page: move MEMS using buttons, operate laser by starting flash lamp, then QS, moving MEMS will first turn off QS, for safety, so needs to be turned on again]]
|}
|}
Figure 9 - MIDAS laser page


<span id="_Ref567783656" class="anchor"><span id="_Toc647854566" class="anchor"></span></span>Figure 8 - MIDAS laser page
=== Sequencer operation ===
To perform calibration runs with the TPC, the operation of the laser and MEMS is handled by the MIDAS sequencer, using the sequence file ''laserRuns.msl'', which can be found in the alphasoft git. It will save the existing trigger settings, then change them to the special laser run settings, and turn on the flash lamp. It will then loop over the 4 fiber ports, and for each one
# move the mirror to aim the laser
# turn on Q-switch
# take a fixed-length data run


=== Trigger setup ===
Afterwards it restores the saved settings.
 
'''NOTE: INTERRUPTING THE SEQUENCER BREAKS THE RESTORE:''' if you interrupt the sequencer for any reason, it will not restore the settings, so the DAQ will not be put back in normal operating condition. This is a shortcoming of the MIDAS sequencer that has been remedied by adding a dedicated AT_EXIT function call in more recent versions of MIDAS, so some time after the DAQ is updated to a newer MIDAS version, this should be fixed.
 
== Trigger setup ==
The laser outputs a positive 5V TTL signal of a considerable length, while the Alpha16 NIM trigger requires an inverted NIM-type signal, i.e. +500mV or so, nanoseconds to tens of nanoseconds in length. We found that the TTL-NIM converter actually only inverts the TTL signal (when switched to COMPL), without changing the length, while a simple NIM discriminator only works on a negative analog signal, not the positive TTL. So the TTL gets inverted by the TTL-NIM, that inverted signal then goes into a discriminator, creating a short NIM-like pulse. Using the complement output gives us the positive instead of negative pulse. (DC offsets should not affect these electronics too much.)
The laser outputs a positive 5V TTL signal of a considerable length, while the Alpha16 NIM trigger requires an inverted NIM-type signal, i.e. +500mV or so, nanoseconds to tens of nanoseconds in length. We found that the TTL-NIM converter actually only inverts the TTL signal (when switched to COMPL), without changing the length, while a simple NIM discriminator only works on a negative analog signal, not the positive TTL. So the TTL gets inverted by the TTL-NIM, that inverted signal then goes into a discriminator, creating a short NIM-like pulse. Using the complement output gives us the positive instead of negative pulse. (DC offsets should not affect these electronics too much.)


A previous attempt resulted in cryptic error messages from the laser, as apparently the discriminator fed some sort of signal back into the laser QS-out. This arrangement seems to avoid that.
A previous attempt resulted in cryptic error messages from the laser, as apparently the discriminator fed some sort of signal back into the laser QS-out. This arrangement seems to avoid that.
<span id="_Fig_trigger" class="anchor"></span>
{|
{|
|
|
[[File:Laser trigger.svg|thumb|Trigger conversion setup, with approximate signal shapes]]
[[File:Laser trigger.svg|thumb|Trigger conversion setup, with approximate signal shapes]]
|}
|}
<span id="_Ref567783657" class="anchor"><span id="_Toc647854567" class="anchor"></span></span>Figure 9 - Laser trigger setup
Figure 10 - Laser trigger setup
 
=== Test procedure ===
To confirm laser light is indeed entering the TPC and causing photoelectrons, the ADC32 grandOr trigger of the TPC DAQ can be made to produce a signal on a LEMO DAC output of an ALPHA16 ADC, it can then be observed on a scope triggered on the laser's Q-switch output signal.consult laser experts
 
<span id="_Fig_tpclasersig" class="anchor"></span>
{|
|
[[File:PXL 20260813 124249603.jpg|thumb|TPC adc32grandOr signal in blue, triggered on Q-switch signal in yellow. The two peaks are photoelectrons liberated from the pads (first peak) and Aluminium strips (second). The distance between the peaks is (almost) the full drift time.]]
|}
Figure 11 - Test for laser light in TPC
 
== Photodiode tests ==
The laserbox also contains a Thorlabs photodiode, which can be used for diagnostic and alignment purposes. It is normally connected to the 5th fiber port in the centre of the array, labeled ''PD'', and can be read out via a BNC port on the outside of the laserbox. To read out the signal, connect the BNC to a scope with 1MOhm impedance and set the scales appropriately to see a signal up to 3V or so with about 1ms duration. Set the trigger level to tens of mV.
 
* For simple quick diagnostic, aim the MEMS at the PD port, then fire the laser and observe the signal
* To diagnose other port positions, move fiber over to port of interest, then do the same
* To diagnose fibers up to the junction box, move PD into junction box and attach short fiber with correct connector to connect to laser-side fibers
* To determine/confirm fiber port positions, turn off laser frontend and operate laser manually, then run executable MEMS_scan (this requires a Tektronix scope to be connected to the pi via USB and have the PD signal in channel 1, call ''MEMS_scan -h'' for parameters)


== Troubleshooting ==
== Troubleshooting ==
Line 250: Line 324:
|-
|-
| DAQ shows double rate (100Hz instead of 50Hz) || Try increasing threshold on discriminator
| DAQ shows double rate (100Hz instead of 50Hz) || Try increasing threshold on discriminator
|-
| No laser light seen in PD or TPC || Check if control LEDs on MEMS driver board are on, specifically blue, if not 30kHz clock to MEMS is not running and MEMS won't move
|-
| Still no laser light || Take required precautions to run laser with open box and check for laser light at laser head with fluorescent card, '''consult laser experts'''
|-
|}
|}


Line 256: Line 335:
To reproduce the T2K laser calibration configuration in the Alpha-g rTPC, it would require shining the laser normal to the inner cathode surface. While the drift region is about 1m long for T2K, it is about 5cm in the Alpha-g. In addition the outer surface of the rTPC is fully occupied with the cathode pad readout electronics boards preventing any acceptable insertion of laser light direction normal to the tangent of its surface.
To reproduce the T2K laser calibration configuration in the Alpha-g rTPC, it would require shining the laser normal to the inner cathode surface. While the drift region is about 1m long for T2K, it is about 5cm in the Alpha-g. In addition the outer surface of the rTPC is fully occupied with the cathode pad readout electronics boards preventing any acceptable insertion of laser light direction normal to the tangent of its surface.


The remaining possible solution for the Alpha-g laser calibration is to shine the laser from its endplates. There the space is also quite restricted but available. In this configuration see Figure 1, the incident light angle varies from about 45 deg to less than 1deg due to the rTPC geometrical proportions. The electron emission efficiency at these incident angles varies substantially.
The remaining possible solution for the Alpha-g laser calibration is to shine the laser from its endplates. There the space is also quite restricted but available. In this configuration (see [[#_Fig_crosssec|Figure 1]]) the incident light angle varies from about 45 deg to less than 1deg due to the rTPC geometrical proportions. The electron emission efficiency at these incident angles varies substantially.


The shallow angle of the laser light on the central cathode cylinder is a concern for the electron emission efficiency. In order to quantify the parameter for the laser source specification, we have gone through multiple tests at UBC where a vacuum chamber with a 266nm laser was available. A dedicated mounting setup within the vacuum chamber permitted to rotate a plane holding our material sample, see Figure 2. The setup consisted in a double plate “capacitor-like” configuration, with one of the electrodes holding the sample and the other collecting the emitted electrons, see Figure 3. The signal was then collected through a dedicated DAQ for signal charge analysis versus incident angle on different material.
The shallow angle of the laser light on the central cathode cylinder is a concern for the electron emission efficiency. In order to quantify the parameter for the laser source specification, we have gone through multiple tests at UBC where a vacuum chamber with a 266nm laser was available. A dedicated mounting setup within the vacuum chamber permitted to rotate a plane holding our material sample (see [[#_Fig_ubcsetup|Figure]]). The setup consisted in a double plate “capacitor-like” configuration, with one of the electrodes holding the sample and the other collecting the emitted electrons, (see [[#_Fig_ubcschem|Figure]]). The signal was then collected through a dedicated DAQ for signal charge analysis versus incident angle on different material.


<span id="_Fig_ubcsetup" class="anchor"></span>
{|
{|
|
|
[[File:imagec9.png|394x300px]]
[[File:imagec9.png|394x300px]]
 
|
<span id="_Ref479322156" class="anchor"><span id="_Toc493167190" class="anchor"></span></span>Figure 10 - UBC setup for laser test
[[File:imagec10.png|298x223px]]
| [[File:imagec10.png|298x223px]]
|}
|}
Figure 12 - UBC setup for laser test


<span id="_Fig_ubcschem" class="anchor"></span>
[[File:imagec11.png|464x347px]]
[[File:imagec11.png|464x347px]]
 
Figure 13 - Schematic of the UBC test configuration
<span id="_Ref479322142" class="anchor"><span id="_Toc493167191" class="anchor"></span></span>Figure 11 - Schematic of the UBC test configuration


The Aluminum and copper have different “work function” values. The results from T2K showed electron extraction from the aluminum to be about 2 pe/mm<sup>2</sup> while for copper around 0.03 pe/mm<sup>2</sup>. This ratio of 70 provides the necessary contrast for strip localization and therefore local drift time calibration. Similar tests at UBC were set up to possibly confirm these numbers. In addition the work function values found in literature fluctuate sufficiently that the selected materials may overlap with a 266nm wavelength. Therefore we added to our test the target composition (Al, Cu, C) as a variable to see if a better suited material can be identified.
The Aluminum and copper have different “work function” values. The results from T2K showed electron extraction from the aluminum to be about 2 pe/mm<sup>2</sup> while for copper around 0.03 pe/mm<sup>2</sup>. This ratio of 70 provides the necessary contrast for strip localization and therefore local drift time calibration. Similar tests at UBC were set up to possibly confirm these numbers. In addition the work function values found in literature fluctuate sufficiently that the selected materials may overlap with a 266nm wavelength. Therefore we added to our test the target composition (Al, Cu, C) as a variable to see if a better suited material can be identified.


In the Alpha-g case, the laser incident angle is grazing the cathode surface. Hence we included in the incident angle of the laser beam on target as a variable as well (see. Figure 4). Our tests did not reproduce the T2K results. That ratio of collected electron to the estimated number of photons on target versus the photon (laser beam) incident angle is not as large as seen in T2K at normal incidence. But we did expect to see a large variation of electron emission as we move away of a normal incidence. A proposed method to recover those “missing” electron due to the grazing incident angle, was to “scratch” the surface of the target (aluminum) to have some normal area to the incident laser light (note that 90deg incident angle is normal to the surface). This is visible with the “Al scratched 2” plot (the Alpha-g incident angle range is between 0.5 and 45deg). The effect of the scratching does increase the emission at grazing angle quite substantially, which is what we were looking for.
In the Alpha-g case, the laser incident angle is grazing the cathode surface. Hence we included in the incident angle of the laser beam on target as a variable as well (see. [[#_Fig_ubcres1|Figure]]). Our tests did not reproduce the T2K results. That ratio of collected electron to the estimated number of photons on target versus the photon (laser beam) incident angle is not as large as seen in T2K at normal incidence. But we did expect to see a large variation of electron emission as we move away of a normal incidence. A proposed method to recover those “missing” electron due to the grazing incident angle, was to “scratch” the surface of the target (aluminum) to have some normal area to the incident laser light (note that 90deg incident angle is normal to the surface). This is visible with the “Al scratched 2” plot (the Alpha-g incident angle range is between 0.5 and 45deg). The effect of the scratching does increase the emission at grazing angle quite substantially, which is what we were looking for.


Unfortunately it was not possible to get reasonable values comparing different target materials. It remains unclear why this is the case.
Unfortunately it was not possible to get reasonable values comparing different target materials. It remains unclear why this is the case.
Line 296: Line 376:
|}
|}


<span id="_Fig_ubcres1" class="anchor"></span>
[[File:imagec12.png|270x197px]][[File:imagec13.png|280x201px]]
[[File:imagec12.png|270x197px]][[File:imagec13.png|280x201px]]
Figure 14 - Photoemission versus incident angle for Aluminum


<span id="_Ref479691369" class="anchor"><span id="_Toc493167192" class="anchor"></span></span>Figure 12 - Photoemission versus incident angle for Aluminum
<span id="_Fig_ubcres2" class="anchor"></span>
 
[[File:imagec14.png|465x342px]]
[[File:imagec14.png|465x342px]]
 
Figure 15 - Ratio of electron per photon versus incident angle for different base material
<span id="_Ref479322103" class="anchor"><span id="_Toc493167193" class="anchor"></span></span>Figure 13 - Ratio of electron per photon versus incident angle for different base material

Latest revision as of 17:46, 2 September 2026

Back to Main Page

Table of Figures

Figure 1 – rTPC cross-section with at the bottom the central cathode with Al strips. A laser plane illuminates the central cathode for electron production

Figure 2 - Inner cylinder with copper cathode and Aluminum strips

Figure 3 - Optical head assembly to be mounted on the endplate

Figure 4 - Model of the laser equipment at the endplate with light simulated plane on first Aluminum strip

Figure 5 - Pulse height vs pad row from prototype

Figure 6 - Laser shutter positions

Figure 7 - Laser fiber connection box

Figure 8 - Laser box and MEMS driver

Figure 9 - MIDAS laser page

Figure 10 - Laser trigger setup

Figure 11 - Test for laser light in TPC

Figure 12 - UBC setup for laser test

Figure 13 - Schematic of the UBC test configuration

Figure 14 - Photoemission versus incident angle for Aluminum

Figure 15 - Ratio of electron per photon versus incident angle for different base material

Purpose

The calibration system for the ALPHA-g rTPC is to provide a reference measurement of the electron drift direction and speed for a known configuration. These parameters are subject to environmental changes (pressure, temperature, gas mixture) as well as external drift voltage and magnetic field.

Method

For comparing to a reference measurement, a consistent number of primary electrons at a defined geometrical location are required. Two methods have been considered.

  • Alpha source (Am241) mounted on the central cathode surface. This technique is well understood and working well in particular for a long drift detector. The emission rate from the calibration source can be adjusted but cannot be controlled. In our case, the impact of the calibration event rate on the overall physics trigger rates is to be considered as once the source are installed, the emission cannot be stopped.
  • Laser beam striking aluminized surface on the central cathode to achieve electron emission. This technique has been used for the T2K TPC detector with success. The work function of the aluminum being lower than the one of copper, an increase of electron emission from the aluminum will be distinguishable relative to the copper. In our case cathode surface is copper with aluminum strips at known locations. Therefore it is possible to “image” the aluminum positions from the emitted drifting electrons to the anodes. The intended laser wavelength is to be the same as in T2K (Nd:YAG frequency-quadrupled to 266nm).

The laser method with its external trigger capability and better defined electron emission has been deemed the preferred solution.

References

Concept

The top and bottom endplates have provisions for inserting the necessary optic module to deliver a laser plane intersecting the inner cathode cylinder. The initial laser beam in the order of 0.5mm in diameter is converted into a plane through a transparent cylindrical quartz rod of 5mm diameter within the detector volume. The plane has an aperture angle in the order of 90deg illuminating both sides of the drift path. Aluminum strips on pre-defined locations along the inner cathode will emit electrons when struck by the 266nm laser light. Using a triggered pulse laser source, the electron arrival on the anode wire and the induced charge on the cathode pads along the anode wire will provide the timing and Z position of the intersection point of the aluminum strip and the laser plane. The light plane is slightly rotated and tilted off the z-axis in order separate the Al strip electron cloud from reaching all the same wires. For the given rotation and tilt angles (45deg, 1deg) we have spaced the Al strips such that 2 anode wires will be separating the 2 electron clouds of two consecutive Al strips.

Figure 1 – rTPC cross-section with at the bottom the central cathode with Al strips. A laser plane illuminates the central cathode for electron production.

Mechanical design

The mechanical modification to the endplate for the laser calibration option is minimal. They consist in:

  • Placing 2 laser access holes in each endplate, giving us the option to feed 4 laser optic modules to the rTPC (see Figure). The advantage of multiple access points is to better cover the whole drift volume and to illuminate both ends of the detector with similar incident angles.
  • Additional 9 Aluminum strips of 6mm width wrapped around the inner cylinder at 9 locations (see Figure). Table 1 list the position of the 9 strips, the final column shows the phi coordinate at which the laser light hits the given strip for one possible set of geometric parameters of the optics package.

Strip number Nominal z [mm] Laser phi in degrees
0 -1060 -17.647
1 -795 -7.450
2 -530 0.447
3 -265 7.225
4 0 13.328
5 265 18.978
6 530 24.310
7 795 29.412
8 1060 34.348

Table 1 - Al strip position

Figure 2 - Inner cylinder with copper cathode and Aluminum strips

Figure 3 - Optical head assembly to be mounted on the endplate 6

Figure 4 - Model of the laser equipment at the endplate with light simulated plane on first Aluminum strip.

Specifications

Parameter Value
Laser Pulsed beam 0..100Hz triggerable
Wavelength 266 nm
Power 2uW
Pulse Energy 2.6mJ
Pulse Width 8.8ns
Near Field Beam Diameter 1.7mm
Divergence at 86.5% 1.8mrad
Aluminum strip width 6mm
Aluminum strip length Full circumference
Aluminum strip pitch in the z-direction 265mm
Number of Aluminum strips 9
Aperture angle 90 deg
Tilt angle 1 deg
Rotation angle 45 deg

Tests of the laser calibration system using the prototype show a clear pattern of the Al strips. Analysis of the pattern reproduces the strip position with good accuracy. Despite the strip pitch in the prototype being much tighter at 40mm instead of 265mm the peaks are very clearly separated in the z-pads. The amplitude variation between peaks is due to the laser intensity distribution which has not been investigated in detail. It is not a high priority because the laser system is not meant for pulse height calibration, but rather for time and position.

Figure 5 - Pulse height vs pad row from prototype

The laser plane is not inclined in the prototype, so a separation on the anode wires is impossible. The wider spacing for the final detector was chosen to allow for a good separation in the anodes.

As mentioned above the laser intensity distribution is not well understood yet, but we are confident we can illuminate the 4-5 strips closest to each laser port. The combination of shallow incidence angle and low geometric efficiency makes it questionable whether enough light can reach the far end of the TPC. It may be possible to increase the laser intensity to reach further strips at the expense of saturating the signals in the intensity maximum.

Physical safety measures and Interlock

Shutter and Box

The laser is equipped with a physical shutter, that prevents laser light from coming out even if the laser is firing. Ensure the shutter is closed before disconnecting fiber. Ensure it's open when trying to operate.

Physical shutter closed, laser is safe
Physical shutter open, laser can operate

Figure 6 - Laser shutter positions

The laser head and multiplexer are> I noticed that for some reason the per-port Q-switch delay setting in the ODB was way higher for port T11 (elog:6574/1 [3], higher delay -> less laser power), and indeed setting it to a value more in line with the others (210) brings back the coincidence triggers, so we do in fact get laser light from all 4 ports. > Running a longer sequence now, and will turn on PWBs later.

enclosed in a locked metal case, which is hooked up to the interlock input on the laser power supply, which means the laser will not fire if the case is open. (As always with electronic interlocks, do not rely on this, but confirm the interlock light on the handheld control box is blinking when the box is open.)

The combination for the lock is the wavelength of the laser.

Fiber protection

The laser fibers are enclosed in a steel conduit, with a small locked metal box protecting the interconnections between laser-side and TPC-side fibers.

Locked box containing the interconnects between laser-side and TPC-side fibers
Interconnects between laser-side and TPC-side fibers

Figure 7 - Laser fiber connection box

MEMS multiplexer

In addition to the laser head itself, the laser box contains a fiber multiplexer based around a micro-electromechanical system (MEMS) mirror from Mirrorcle Technologies:

Device Part No. documentation
MEMS mirror A7M20.2-2000AL File:MirrorcleTech Datasheet A7M20.2-2000AL.pdf
Driver board DR-10-056-00 File:Mirrorcle MEMS Drivers 5.x - User Guide.pdf
Breakout board BRK-DRIVER-5.x

The MEMS driver board is controlled via SPI from a raspberry pi, which also provides a required 30kHz clock and the 5V supply voltage it needs. The driver board generates about 200V steering voltage for the MEMS from this supply, so should not be touched when operational. When the driver board is connected correctly to a correctly configured raspberry pi, both the green and blue LED should be illuminated. The red LED is only lit when the high voltage is turned on, which is done via a GPIO pin from the pi.

Laserbox interior with head (A), lens (B), MEMS mirror (C), MEMS driver (D), fiber port array (E)
MEMS driver board with controller connector (A, to breakout board, then rPi), status LEDs (B), voltage testpoints (C), HV connector to MEMS (D)

Figure 8 - Laser box and MEMS driver board

Operation

Manual operation / Switching on

The laser gets turned on with the key at the front of the power supply unit. After turning on it may take a while (~5min) to warm up the cooling water (interlock message is displayed on the handheld control unit). Other interlock messages indicate something is wrong.

Once the system is started up it can be operated from the handheld control unit, by pressing first the flash start, then after a mandatory waiting period the QS start buttons. This manual operation should not be necessary.

Midas operation

The Midas laser page mirrors the important parts of the manual control. In order to use it the feLaser frontend must be running. To start firing the laser, press the "Start flash lamp" button, then wait until the status line shows "flash", then press "Start Q-Switch". Pausing the laser briefly can be done by just stopping the Q-Switch, for longer breaks turn off the flash lamp. (Turning off the flash lamp automatically turns off the Q-Switch, so there is no need to turn it off manually.

Midas laser page: move MEMS using buttons, operate laser by starting flash lamp, then QS, moving MEMS will first turn off QS, for safety, so needs to be turned on again

Figure 9 - MIDAS laser page

Sequencer operation

To perform calibration runs with the TPC, the operation of the laser and MEMS is handled by the MIDAS sequencer, using the sequence file laserRuns.msl, which can be found in the alphasoft git. It will save the existing trigger settings, then change them to the special laser run settings, and turn on the flash lamp. It will then loop over the 4 fiber ports, and for each one

  1. move the mirror to aim the laser
  2. turn on Q-switch
  3. take a fixed-length data run

Afterwards it restores the saved settings.

NOTE: INTERRUPTING THE SEQUENCER BREAKS THE RESTORE: if you interrupt the sequencer for any reason, it will not restore the settings, so the DAQ will not be put back in normal operating condition. This is a shortcoming of the MIDAS sequencer that has been remedied by adding a dedicated AT_EXIT function call in more recent versions of MIDAS, so some time after the DAQ is updated to a newer MIDAS version, this should be fixed.

Trigger setup

The laser outputs a positive 5V TTL signal of a considerable length, while the Alpha16 NIM trigger requires an inverted NIM-type signal, i.e. +500mV or so, nanoseconds to tens of nanoseconds in length. We found that the TTL-NIM converter actually only inverts the TTL signal (when switched to COMPL), without changing the length, while a simple NIM discriminator only works on a negative analog signal, not the positive TTL. So the TTL gets inverted by the TTL-NIM, that inverted signal then goes into a discriminator, creating a short NIM-like pulse. Using the complement output gives us the positive instead of negative pulse. (DC offsets should not affect these electronics too much.)

A previous attempt resulted in cryptic error messages from the laser, as apparently the discriminator fed some sort of signal back into the laser QS-out. This arrangement seems to avoid that.

Trigger conversion setup, with approximate signal shapes

Figure 10 - Laser trigger setup

Test procedure

To confirm laser light is indeed entering the TPC and causing photoelectrons, the ADC32 grandOr trigger of the TPC DAQ can be made to produce a signal on a LEMO DAC output of an ALPHA16 ADC, it can then be observed on a scope triggered on the laser's Q-switch output signal.consult laser experts

TPC adc32grandOr signal in blue, triggered on Q-switch signal in yellow. The two peaks are photoelectrons liberated from the pads (first peak) and Aluminium strips (second). The distance between the peaks is (almost) the full drift time.

Figure 11 - Test for laser light in TPC

Photodiode tests

The laserbox also contains a Thorlabs photodiode, which can be used for diagnostic and alignment purposes. It is normally connected to the 5th fiber port in the centre of the array, labeled PD, and can be read out via a BNC port on the outside of the laserbox. To read out the signal, connect the BNC to a scope with 1MOhm impedance and set the scales appropriately to see a signal up to 3V or so with about 1ms duration. Set the trigger level to tens of mV.

  • For simple quick diagnostic, aim the MEMS at the PD port, then fire the laser and observe the signal
  • To diagnose other port positions, move fiber over to port of interest, then do the same
  • To diagnose fibers up to the junction box, move PD into junction box and attach short fiber with correct connector to connect to laser-side fibers
  • To determine/confirm fiber port positions, turn off laser frontend and operate laser manually, then run executable MEMS_scan (this requires a Tektronix scope to be connected to the pi via USB and have the PD signal in channel 1, call MEMS_scan -h for parameters)

Troubleshooting

Problem Possible solution
feLaser doesn't start Check messages if both ICE450 and MVAT were found, if not check USB-serial connections
If everything seems connected, on the handheld control unit, navigate to System and switch Serial Link on
BNC interlock Make sure laser enclosure and fiber cover are screwed down correctly.
Laser shows PSU error Disconnect cable from QS out, if error goes away, re-think/re-build trigger system
Other interlocks / errors Refer to manual File:Brio ICE450 Power Supply Manual.pdf
DAQ doesn't trigger Confirm NIM mask is correct in ODB, and TrigEsataNimGrandOr is selected as trigger source (DAQ manual)
DAQ shows low rate Try adjusting pulse width on discriminator
DAQ shows double rate (100Hz instead of 50Hz) Try increasing threshold on discriminator
No laser light seen in PD or TPC Check if control LEDs on MEMS driver board are on, specifically blue, if not 30kHz clock to MEMS is not running and MEMS won't move
Still no laser light Take required precautions to run laser with open box and check for laser light at laser head with fluorescent card, consult laser experts

Appendix I – Bench tests

To reproduce the T2K laser calibration configuration in the Alpha-g rTPC, it would require shining the laser normal to the inner cathode surface. While the drift region is about 1m long for T2K, it is about 5cm in the Alpha-g. In addition the outer surface of the rTPC is fully occupied with the cathode pad readout electronics boards preventing any acceptable insertion of laser light direction normal to the tangent of its surface.

The remaining possible solution for the Alpha-g laser calibration is to shine the laser from its endplates. There the space is also quite restricted but available. In this configuration (see Figure 1) the incident light angle varies from about 45 deg to less than 1deg due to the rTPC geometrical proportions. The electron emission efficiency at these incident angles varies substantially.

The shallow angle of the laser light on the central cathode cylinder is a concern for the electron emission efficiency. In order to quantify the parameter for the laser source specification, we have gone through multiple tests at UBC where a vacuum chamber with a 266nm laser was available. A dedicated mounting setup within the vacuum chamber permitted to rotate a plane holding our material sample (see Figure). The setup consisted in a double plate “capacitor-like” configuration, with one of the electrodes holding the sample and the other collecting the emitted electrons, (see Figure). The signal was then collected through a dedicated DAQ for signal charge analysis versus incident angle on different material.

Figure 12 - UBC setup for laser test

Figure 13 - Schematic of the UBC test configuration

The Aluminum and copper have different “work function” values. The results from T2K showed electron extraction from the aluminum to be about 2 pe/mm2 while for copper around 0.03 pe/mm2. This ratio of 70 provides the necessary contrast for strip localization and therefore local drift time calibration. Similar tests at UBC were set up to possibly confirm these numbers. In addition the work function values found in literature fluctuate sufficiently that the selected materials may overlap with a 266nm wavelength. Therefore we added to our test the target composition (Al, Cu, C) as a variable to see if a better suited material can be identified.

In the Alpha-g case, the laser incident angle is grazing the cathode surface. Hence we included in the incident angle of the laser beam on target as a variable as well (see. Figure). Our tests did not reproduce the T2K results. That ratio of collected electron to the estimated number of photons on target versus the photon (laser beam) incident angle is not as large as seen in T2K at normal incidence. But we did expect to see a large variation of electron emission as we move away of a normal incidence. A proposed method to recover those “missing” electron due to the grazing incident angle, was to “scratch” the surface of the target (aluminum) to have some normal area to the incident laser light (note that 90deg incident angle is normal to the surface). This is visible with the “Al scratched 2” plot (the Alpha-g incident angle range is between 0.5 and 45deg). The effect of the scratching does increase the emission at grazing angle quite substantially, which is what we were looking for.

Unfortunately it was not possible to get reasonable values comparing different target materials. It remains unclear why this is the case.

The material options for strip/cathode combinations were aluminum on copper or aluminum on carbon loaded Kapton. Both cathode materials are suitable for the rTPC. In light of the prototype results shown above it was decided to stick with the copper cathode and scratched aluminum strips.

Wavelength 266nm, Photon energy: 4.66 eV

Material Work function (eV)
Cu 4.53 – 5.10
Al 4.06 – 4.26
C ~5

Figure 14 - Photoemission versus incident angle for Aluminum

Figure 15 - Ratio of electron per photon versus incident angle for different base material