DS-VX

From DaqWiki
Jump to navigation Jump to search

Darkside Digitizer: CAEN VX2745

Connecting

Each digitizer has a webpage served at its local IP address, port 80. You can point your browser to this page after connecting.

Connect to the vx in one of two ways:

  1. Ethernet: if the module is setup with DHCP and a server has been configured.
  2. USB: the module is configured to run as ethernet-over-usb. Download the driver from the downloads page or get version 1.3 here. Then do the following from inside that extracted file:
sudo ./install.sh 
sudo ./regPID.sh

The former adds the IP address to your /etc/hosts file, which then allows access.

Recovery Mode

To go into recovery mode, power down the board, and apply a jumper between the TX and RX terminals (as per this elog). You can then point your browser to http://[fda7:cae0::1]/ to reset the firmware. There is no need to re-run regPID.sh.

Firmware Updates

Firmware is in an SVN repository (ask Yair for access) and is custom: contains some parts CAEN and some parts specific to Darkside. As of this writing the current firmware file is V2745-OpenDPP-2026102907.cup.

Firmware updates are accomplished via the webpage: Firmware (left panel) -> Upload new firmware (blue button).

Firmware updates will also update the webpage.

Note for 2026102907 firmware: the "Apply it after upload" is faulty (elog). Here are the steps to take instead (hopefully this is fixed in future versions):

  1. Upload the firmware to the board, without applying it (Big blue button)
  2. Mark it as default, without applying it (blue "..." -> More actions -> set as default)
  3. Reboot the board (top right username -> reboot device)

If the firmware update goes drastically wrong, go to recovery mode.

Software Updates

Software is on bitbucket: https://bitbucket.org/ttriumfdaq/dsproto_vx2740/src/develop/

Go to the setup section of the README for installation notes.

Notable Elog Entries