Documentation for the Central Cathode High Voltage control through MSCB
The GPIB/MSCB interface provides monitoring and control to the BERTAN HV power supplies.
The Bertan 30KV unit is controlled through the GPIB port. The GPIB address is set to 4 by default. The main MSCB variables are:
0: Status 8bit U 32 (0x20/00100000) byte
Ramp up state 0x01
Ramp down state 0x02
Voltage limit reached 0x04
Current limit reached 0x08
Low current state 0x10
High current state 0x20
1: Udemand 32bit F 0 volt
2: Umeas 32bit F 0 volt
3: Imeas 32bit F 0 ampere
4: RampUp 16bit U 50 (0x0032) volt
5: RampDown 16bit U 50 (0x0032) volt
6: Ulimit 32bit F 28000 volt
7: Ilimit 32bit F 75 microampere
8: warning STR32 "Warning! High current - x.xxxx"
SS level in the TPC rack. 3 BERTAN units with corresponding ccbertan MSCB interface. Fourth one on the top of the BERTAN-TPC3