==Introduction ==
==Introduction ==
Line 22: Line 15:
    
    
== EPICS switch ==
== EPICS switch ==
The EPICS parameter that controls switching is called "BNMR:BNQRSW:ENBSWITCHING". Switching must be
The EPICS parameter that controls switching is called <code>BNMR:BNQRSW:ENBSWITCHING</code>. Switching must be
* '''enabled''' for dual channel mode
* '''enabled''' for dual channel mode
* '''disabled''' for single channel mode  
* '''disabled''' for single channel mode  
Line 32: Line 25:
[[Image:bnmr_bnqr_switching_en.png|frame|center|Figure 1: EPICS BNMR-BNQR Switching set for dual channel mode]]
[[Image:bnmr_bnqr_switching_en.png|frame|center|Figure 1: EPICS BNMR-BNQR Switching set for dual channel mode]]


Experimenters must also ensure that the DAQ ODB parameter {{Odbpath|path=Enable dual channel mode}} is set to the appropriate value. The ODB parameter {{Odbpath|path=enable epics switch checks}} (on the custom status page) should also be set true (except for testing). This causes the MIDAS client [[BNMR rf_config]] to compare the EPICS switch position with the value of {{Odbpath|path=Enable dual channel mode}}. If the switch is not set correctly,  {{Client|name=rf_config}} prevents the run from starting.
Experimenters must ensure that the DAQ ODB parameter {{Odbpath|path=/Scanning/BNXRCustomCycleLogic/Settings/Enable dual channel mode}} (called <code>Enable dual channel mode</code> on the [[BNMR: Custom Status page|custom status page]]) is set appropriately.
 
== Frontend sanity-checks ==
 
Unless performing tests, the ODB parameter {{Odbpath|path=/Scanning/BNXRCustomCycleLogic/Settings/Enable EPICS access checks}} (called <code>Check switching in run start</code> on the [[BNMR: Custom Status page|custom status page]]) should be set true. This causes the [[BNMR: frontend|frontend]] to validate the beam switching setup:
* Ensures the DAQ {{Odbpath|path=Enable dual channel mode}} flag must match the EPICS <code>BNMR:BNQRSW:ENBSWITCHING</code> flag.
* Ensures the EPICS <code>ILE2:BIAS15:VOL.ASG</code> value must be set to <code>expbnmr</code> (for both {{bnmqr|join=and}}).
* Ensures the EPICS <code>BNMR:BNQRSW:SCHEDULE.SCAN</code> value is set to <code>.1 second</code>. If this value is <code>Disabled</code>, then switching never happens and the DAQ does not run correctly.


== Helicity ==
== Helicity ==
Line 43: Line 43:
In dual channel mode, the PPG Cycle is started by an external signal (PPG Start/Beam Ready) that occurs at the end of the BNMR or BNQR Delay (Figure 2).  Usually the PPG cycle is divided into 3 regions: prebeam, beam on , and post-beam. Once the PPG cycle is started, the PPG Output "Beam On" controls when and how long the beam is on. The timing of these signals is controlled by the [[PPG Parameters]].
In dual channel mode, the PPG Cycle is started by an external signal (PPG Start/Beam Ready) that occurs at the end of the BNMR or BNQR Delay (Figure 2).  Usually the PPG cycle is divided into 3 regions: prebeam, beam on , and post-beam. Once the PPG cycle is started, the PPG Output "Beam On" controls when and how long the beam is on. The timing of these signals is controlled by the [[PPG Parameters]].
    
    
The experimenter must be careful to set the timing of the BNMR and BNQR cycles such that the Beam On period of the PPG Cycle has finished before the beam is switched to the other channel. The post-beam period of one channel generally overlaps part of the cycle of the other channel. This timing is controlled by the EPICS parameters ''BNMR Period (s), BNQR Period (s), BNMR Delay (s), BNQR Delay (s)''. These are available on the EPICS Kicker Control/BNMR-BNQR Switching page (Figure 1). The BNMR and BNQR Delays must be set to give the Helicity time to change. They are equivalent to the ODB parameter {{Odbpath|path=Helicity Sleep time}} in Single Channel mode. This parameter is ignored in dual channel mode.
The experimenter must be careful to set the timing of the BNMR and BNQR cycles such that the Beam On period of the PPG Cycle has finished before the beam is switched to the other channel. The post-beam period of one channel generally overlaps part of the cycle of the other channel. This timing is controlled by the EPICS parameters ''BNMR Period (s), BNQR Period (s), BNMR Delay (s), BNQR Delay (s)''. These are available on the EPICS Kicker Control/BNMR-BNQR Switching page (Figure 1). The BNMR and BNQR Delays must be set to give the Helicity time to change. They are equivalent to the ODB parameter {{Odbpath|path=Helicity Sleep time}} in Single Channel mode. This parameter is ignored in dual channel mode.
 


== Single Channel Mode ==
== Single Channel Mode ==
In '''single channel mode''', the beam is directed to '''one''' of the channels ({{bnmqr|join=or}}). The [[#EPICS switch]] and the DAQ ODB parameter {{Odbpath|path=Enable dual channel mode}} must both be set to "Disabled".
In '''single channel mode''', the beam is directed to '''one''' of the channels ({{bnmqr|join=or}}). The [[#EPICS switch]] and the DAQ ODB parameter {{Odbpath|path=/Scanning/BNXRCustomCycleLogic/Settings/Dual channel mode}} must both be set to off/false/disabled.


'''Only the DAQ system for the channel with the beam should be running.'''  The other DAQ system must not be run (except by "experts" for [[#Testing]] under certain conditions). This is because there is currently NO CHECK against both DAQ systems running simultaneously in single channel mode, and both trying to flip the helicity. This would ruin the data for the active channel with the beam.  
'''Only the DAQ system for the channel with the beam should be running.'''  The other DAQ system must not be run (except by "experts" for [[#Testing]] under certain conditions). This is because there is currently NO CHECK against both DAQ systems running simultaneously in single channel mode, and both trying to flip the helicity. This would ruin the data for the active channel with the beam.  


In single channel mode, the PPG Cycle is started by a software command given by the  [[BNMR: Frontend|frontend]].  
In single channel mode, the PPG Cycle is started by a software command given by the  [[BNMR: frontend|frontend]].  


== Testing ==
== Testing ==

Revision as of 11:39, 27 April 2022

Introduction

dual channel mode
is when the beam is switched alternately between the two experiments bnmr and bnqr .
single channel mode
is when no switching of the beam takes place. and the beam is sent to one of the two experiments (bnmr or bnqr), that experiment is said to be running in #Single Channel Mode.

Switching the beam alternately between the two experiments is done by the Controls Group using the #EPICS switch.

Currently, only Type 2 (TD) experiments can be run in dual channel mode. When running in dual channel mode, both the DAQ systems bnmr and bnqr may be running and taking data, therefore dual channel mode makes better use of the beam time.

In #Single channel mode, only one DAQ system (bnmr or bnqr) is active.

EPICS switch

The EPICS parameter that controls switching is called BNMR:BNQRSW:ENBSWITCHING. Switching must be

  • enabled for dual channel mode
  • disabled for single channel mode

Only the Controls Group can change this switch. The experimenters telephone the Control Room and ask them set this switch as required to run in dual channel or single channel mode. Figure 1 shows the switching set for Dual channel mode (scheduled switching enabled, manual switching disabled). This page can be found on the EPICS page display by clicking on the button "beam line switch" on the page "ILE2A3/BNMR OPTICS".

It is important to match the timing of the beam switch (BNMR Period/Delay and BNQR Period/Delay) to the PPG Cycle of each experiment (see #Timing).

Figure 1: EPICS BNMR-BNQR Switching set for dual channel mode

Experimenters must ensure that the DAQ ODB parameter /Scanning/BNXRCustomCycleLogic/Settings/Enable dual channel mode (called Enable dual channel mode on the custom status page) is set appropriately.

Frontend sanity-checks

Unless performing tests, the ODB parameter /Scanning/BNXRCustomCycleLogic/Settings/Enable EPICS access checks (called Check switching in run start on the custom status page) should be set true. This causes the frontend to validate the beam switching setup:

  • Ensures the DAQ Enable dual channel mode flag must match the EPICS BNMR:BNQRSW:ENBSWITCHING flag.
  • Ensures the EPICS ILE2:BIAS15:VOL.ASG value must be set to expbnmr (for both bnmr and bnqr).
  • Ensures the EPICS BNMR:BNQRSW:SCHEDULE.SCAN value is set to .1 second. If this value is Disabled, then switching never happens and the DAQ does not run correctly.

Helicity

The PPG requests the helicity it requires for the next cycle by a signal from the PPG Output "Pol" connected to the helicity switch box. In Single Channel mode, the helicity is immediately set to this value (UP/DOWN). In Dual Channel mode, the helicity state is not changed until that channel is active (see #Timing).

Timing

Figure 2 shows a timing diagram for dual channel mode.

Figure 2: Timing diagram for Dual Channel mode

In dual channel mode, the PPG Cycle is started by an external signal (PPG Start/Beam Ready) that occurs at the end of the BNMR or BNQR Delay (Figure 2). Usually the PPG cycle is divided into 3 regions: prebeam, beam on , and post-beam. Once the PPG cycle is started, the PPG Output "Beam On" controls when and how long the beam is on. The timing of these signals is controlled by the PPG Parameters.

The experimenter must be careful to set the timing of the BNMR and BNQR cycles such that the Beam On period of the PPG Cycle has finished before the beam is switched to the other channel. The post-beam period of one channel generally overlaps part of the cycle of the other channel. This timing is controlled by the EPICS parameters BNMR Period (s), BNQR Period (s), BNMR Delay (s), BNQR Delay (s). These are available on the EPICS Kicker Control/BNMR-BNQR Switching page (Figure 1). The BNMR and BNQR Delays must be set to give the Helicity time to change. They are equivalent to the ODB parameter Helicity Sleep time in Single Channel mode. This parameter is ignored in dual channel mode.

Single Channel Mode

In single channel mode, the beam is directed to one of the channels (bnmr or bnqr). The #EPICS switch and the DAQ ODB parameter /Scanning/BNXRCustomCycleLogic/Settings/Dual channel mode must both be set to off/false/disabled.

Only the DAQ system for the channel with the beam should be running. The other DAQ system must not be run (except by "experts" for #Testing under certain conditions). This is because there is currently NO CHECK against both DAQ systems running simultaneously in single channel mode, and both trying to flip the helicity. This would ruin the data for the active channel with the beam.

In single channel mode, the PPG Cycle is started by a software command given by the frontend.

Testing

Both DAQ systems may be run in single channel mode simultaneously for testing under certain conditions. Outside the beam period, it is sufficient to disable helicity flipping on one of the DAQ systems, so that for example BNMR flips the helicity, and BNQR does not. In that case, the Helicity check must be disabled on BNQR.

Alternatively, and especially during the beam period, the beam on/off and helicity outputs from the PPG on the test system should be disconnected so that the test DAQ cannot interfere with the data taking run. In this case, helicity flipping can be enabled on the disconnected system (but helicity is not actually flipped), but Helicity check must be disabled.