Entry  23 Mar 2020, Ivo Schulthess, Forum, Save data to FTP 
Dear all
I try to save data to an FTP server but don't get any data on the server. Midas does not complain or message any error but also nothing gets saved. Does
somebody have experience with this? I use the following settings for the ODB mlogger channel settings: Type: FTP, Filename: server.com, 21, user, pw, .,
    Reply  24 Mar 2020, Ivo Schulthess, Forum, Save data to FTP 
> > I try to save data to an FTP server but don't get any data on the server. Midas does not complain or message any error but also nothing gets saved.
Does somebody have experience with this? I use the following settings for the ODB mlogger channel settings: Type: FTP, Filename: server.com, 21, user,
pw, ., run%06d.mid, Format: MIDAS, Output: FILE. What would be the Output: FTP setting for? I tried this but it does not work at all. 
    Reply  24 Mar 2020, Ivo Schulthess, Forum, Save data to FTP 
> Logging directly from the midas logger to FTP is a bit cumbersome. In case of delays during login etc. this can throttle the whole DAQ chain. 
> What we use in our lab is to write to local disk, then use the lazylogger (https://midas.triumf.ca/MidasWiki/index.php/Lazylogger) to copy the 
> local files to a remote FTP server. This way we de-couple data taking from backup, making the system much more swift.
Entry  07 Apr 2020, Ivo Schulthess, Suggestion, Sequencer loop break 
I am using the Midas sequencer to run subsequent measurements in a loop, without 
knowing how many iterations in advance. Therefore, I am using the "infinity" 
option. Since I have other commands after the loop, it would be nice to have the 
    Reply  23 Apr 2020, Ivo Schulthess, Suggestion, Sequencer loop break 
> You can do that with the "GOTO" statement, jumping to the first line after the loop.

> Here is a working example:
Entry  10 Jun 2020, Ivo Schulthess, Forum, slow-control equipment crashes when running multi-threaded on a remote machine 
Dear all

To reduce the time needed by Midas between runs, we want to change some of our periodic equipment to multi-threaded slow-control equipment. To do that
    Reply  12 Jun 2020, Ivo Schulthess, Forum, slow-control equipment crashes when running multi-threaded on a remote machine 
Thanks you two once again for the very fast answers. I tested the example on the local machine and it works perfectly fine. In the meantime I also created
two new drivers for our devices 
and everything works with them, the improvement in time is significant and I will create drivers for all our devices where possible. If they are in a working
Entry  10 Aug 2020, Ivo Schulthess, Bug Report, data missing in runXXXXXX.mid 
Dear all

We just started our beam time at ILL and just found yesterday that for certain 
    Reply  10 Aug 2020, Ivo Schulthess, Bug Report, data missing in runXXXXXX.mid 
> > Dear all
> > 
> > We just started our beam time at ILL and just found yesterday that for certain 
    Reply  10 Aug 2020, Ivo Schulthess, Bug Report, data missing in runXXXXXX.mid 
> Then I'm running out of ideas. Things I would check:

> - Are the file sizes about the same? 
    Reply  10 Aug 2020, Ivo Schulthess, Bug Report, data missing in runXXXXXX.mid 
> with "dump" I meant a true object dump like "hexdump -C run000001.mid". I produced a file with ADC0 and TDC0 
> banks (that's the example from the distribution under exampels/experiments/frontend.cxx), and I get
    Reply  10 Aug 2020, Ivo Schulthess, Bug Report, data missing in runXXXXXX.mid 
> Have you tried longer files? Maybe a few 100 MB or so. Maybe a buffer is not flushed correctly at the end of a run.

Yes, I did. This 7 KB of the data bank is about the limit. If we go only 1 KB higher it seems that we save all data. In 
    Reply  11 Aug 2020, Ivo Schulthess, Bug Report, data missing in runXXXXXX.mid 
> It would be good to pin point there the data is lost. This is the sequence:

> frontend user code -> mfe.c code -> SYSTEM buffer -> mlogger -> disk
    Reply  23 Mar 2022, Ivo Schulthess, Bug Fix, fix for event buffer corruption in bm_flush_cache() 
Thanks for the investigation. Back in 2020, we had some issues of losing data between the system buffer and the logger writing them to disk (https://daq00.triumf.ca/elog-midas/Midas/1966).
This was polled equipment but we had a multithreaded FE running at the same time. Could this be related to the same problem?
Entry  13 Nov 2023, Ivo Schulthess, Forum, mlogger does not HAVE_ROOT 
Good evening, 

I am setting up Midas (v2.1) for a new experiment. We want to save the data in the ROOT format. We installed ROOT from source (v6.28/06), and ROOTSYS is
    Reply  14 Nov 2023, Ivo Schulthess, Forum, mlogger does not HAVE_ROOT 
> Stefan is right. I forgot this. As solution to our troubles, mlogger is built without root support. use rmlogger instead.

> K.O.
    Reply  15 Nov 2023, Ivo Schulthess, Forum, mlogger does not HAVE_ROOT 
> No, I'm not aware of this problem, but I suspect that your events somehow got corrupted. You can try the mdump utility
> or the "Event Dump" web page to peek into your events, maybe you see an issue there. To give you more detailed information,
> I would have to reproduce your problem, which is probably hard without your hardware.
Entry  21 Nov 2023, Ivo Schulthess, Forum, Polled frontend writes data to ODB without RO_ODB 
Good morning, 

In our setup, we have a neutron detector that creates up to 16 MB of polled (EQ_POLLED) data in one event (event limit = 1) that we do not want to have
Entry  19 Jun 2020, Isaac Labrie-Boulay, Info, Building/running a Frontend Task 
To build a frontend task, the user code and system code are compiled and linked 
together with the required libraries, by running a Makefile (e.g. 
../midas/examples/experiment/Makefile in the MIDAS package).
Entry  09 Jun 2020, Isaac Labrie Boulay, Info, Preparing the VME hardware - VME address jumpers. VME_address_jumpers_broken_link.PNG
Hey folks,

I'm currently working on setting up a MIDAS experiment and I am following the 
ELOG V3.1.4-2e1708b5